Table linearization points – KERN & SOHN CCS Version 2.0 User Manual

Page 31

Advertising
background image

CFS/CCS-BA-e-1220

31

 If necessary, at balance zero display using

select the weighing unit [kg or lb], which shall be
used for linearisation. The appeared [] displays
the current weighing unit.

Acknowledge with

.

 Ensure that there are no objects on the weighing

pan. After successful standstill control press the

button.

 The weighing value of the first weight will be

displayed.

 Place weight 1 carefully in the centre of the

weighing plate, wait for the stability display and

confirm by

.

(example)

 The weighing value of the second weight will be

displayed.

 Remove weight 1
 Place weight 2 carefully in the centre of the

weighing plate, wait for the stability display and

confirm by

.

(example)

 After the adjustment the balance will carry out a

self-test. Remove adjusting weight during selftest,
balance will return into weighing mode
automatically.
In case of fault or a wrong weight an error message
(

) will appear, then repeat the

linearization procedure.


8.1.1 Table Linearization points


Max

Display

Weight 1

Weight

Display

Weight 2

Weight

6 kg

„A 2 kg“

2 kg

„A 6 kg“

6 kg

15 kg

„A 5 kg“

5 kg

„A 15 kg“

15 kg

30 kg

„A 10 kg“

10 kg

„A 30 kg“

30 kg

Advertising