9 Basic Operation

9.1 Switching on/off

 to switch-on actuate forward the switch-on/switch-out on the right lower side of

the balance. The balance will carry out a self-test As soon as the weight display
appears, the balance is ready for weighing.

 To switch off press the on/off switch on the right bottom side of the balance at

the back in the CFS 50K-3 models also press forwards.

9.2 Zeroing

Resetting to zero corrects the influence of light soiling on the weighing plate. The
resetting range of the balance was adjusted to ± 2% max. by factory.
Further adjustments are possible in the menu, see chapter 13.

When using as counting system the zeroing range of both balances can be set in
the menu, see chapter 13.

Manual

 Unload the balance

 Press the ZERO button, the balance begins to reset to zero.

The [] symbol over a appears.


Automatic

In the menu the automatic zero point correction can be switched off or the amount
can be changed, see chapter 13.

9.3 Switch over balance/weighing bridge


For piece counting a weighing bridge can be connected via the second balance
interface. In the counting system KERN CCS the quantity counting is made on the
bulk material balance KERN KFP. The reference balance KERN CFS permits due
to its high resolution a very precise determination of the average piece weight.

The second balance can be operated in the same manner like the first balance.