Numeric input of the average piece weight, Automatic reference optimization – KERN & SOHN CCS Version 2.0 User Manual

Page 46

Advertising
background image

46

CFS/CCS-BA-e-1220

11.2 Numeric input of the average piece weight

Set reference

 Input known average unit weight with numerical keys and confirm within 5 secs.

with

or

(CFS 50K-3 models).


If in the weight display as weighing unit [kg] is active, the average piece weight
will be displayed in [g]. If as weighing unit [lb] is active, the average piece weight
is also displayed in [lb].

Count the items

 Tare if necessary, place weighing good and read off the number of items.


When an optional printer is connected, the displayed value can be edited by
pressing (menu settings „P mode“  „Print“  „Au OFF“, chap. 13.2).

Delete unit weight

 Press


11.3 Automatic reference optimization

If at the reference determination the placed weight or the placed piece number is
too small, in the display of the average piece weight the triangle symbol will appear

over

[

]

or

[

]

.

To optimize the calculated average piece weight automatically, add further parts
whose number is smaller than that of the first reference determination. After the
reference optimization sounds a signal tone. For each reference optimization the
average piece weight is newly calculated. As the additional pieces increase the
base for the calculation, the reference also becomes more exact.

By pressing

or

(CFS 50K-3 models) recalculation can be avoided and

thus the reference weight locked.

Automatic reference optimization will be deactivated as soon as the number of
added parts exceeds the saved reference quantity.