15 Second balance interface

When using as counting system, the weighing bridge must be connected with a
suitable cable via the second-balance interface.

9 pin d-subminiature bushing of the
balance

Connection of weighing bridge KERN
KFP

Pin no.:

Balance connection

Pin 1 or 2

EXC+ (5V)

see labelling of load cell

Pin 4 or 5

EXC- (0)

Pin 7

SIG

Pin 8

SIG+

16 RS 232C interface

The balance is typically equipped with a RS 232C interface. The weighing data can
be output depending on the setting in the menu either automatically or by pressing

(or

in CFS 50K-3) via the interface.

This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication
between the weighing balance and the printer.

Use a suitable cable to connect the weighing balance to the interface of the

printer. Faultless operation requires an adequate KERN interface cable.

Communication parameters (baud rate, bits and parity) of balance and

printer must match. Detailed description of the interface parameters see
chap. 13.2, Menu block "

".

16.1 Pin allocation of balance output bushing:


CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 models:



Pin 2: TXD - Output
Pin 3: RXD - Input
Pin 5: GND - Signal ground