7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.

Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing pan.

Information about test weights you will find in the internet under

http://www.kern-sohn.com

In order to avoid errors at the piece quantity determination, both balances

must be adjusted with the same acceleration due to gravity.
In case of non-compliance counting errors will result!

7.1 Adjustment reference balance CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS

30K0.5

Operation

Display

 Switch-on balance and during the selftest press

.

 Use the numeric keys to enter password:

Input the standard password „0000“

 Confirm input by

.

 When using as counting system, the bulk material

balance as well as the reference balance must be
adjusted. The adjustment process must be carried
out on both balances.

Select bulk material or reference balance via

.

The appeared [] displays the current balance.

Acknowledge with

.