□ Calibrate the scale, if It is moved to another location or the environment

'has changed.

□ It Is not necessary to set the gravity acceleration correction, when calibrating

the scale with a calibration mass at the place where the scale is used.

□ Enter the stable weighing data while the MD mark is turned off. If

unstable data Is used, it may cause a weighing error. Arrange the

condition using the FOO filter function.

□ The span calibration needs the zero calibration data. We recommend that

you perform the span calibration Immediately after the zero calibration.

□ Perform the digital linearization function immediately after the zero

calibration. And perform the span calibration continuously.

□ Select a decimal point and weighing unit at the CF function CFOO, CF 01.

□ If you use, the dual or triple range function of the multi-interval scale,

perform the “Range Function", “Zero Calibration" and "Span Calibration".

5.2. Calibration Procedure

5.2.1.

The Whole Procedure

Essential Items

Step A Preparation for calibration.

Step B Minimum division.

Step C Maximum capacity.

Step

B Zero

calibration.

Step G Span calibration

Optional Items

Step D Weighing range function,

i Perform this function just

before zero calibration and

span calibration.

Step F Digital linearization function.

Perform this function after the

zero calibration immediately.

And perform the span

calibration continuously.

Step H Gravity compensation function.

Perform this function after span calibration.

5. Calibration

Page 12

AD-4329 Weighing Indicator