Preparation, Entering the calibration mode, Turn the display on for 10 miniuies – A&D Weighing Indicator AD-4329 User Manual

Page 15: Calibration, Span calibration

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

5.2.2.

Preparation

step 1 Keep the following conditions to calibrate the scale (indicator) correctly.

□ Maintain a constant temperature, stable power and stable input voltage from the

load ceil.

□ Avoid direct sunshine and in the neighborhood of an air conditioner.

□ Do not install the scale (indicator) where there is a strong magnetic field.

Step 2 Turn the display on during 10 minutes.

5.2.3.

Entering the Calibration Mode

Step 3 Press the ICAlj key In the norma! weighing mode. The

two seconds when entering the calibration mode.

CHL

(O is displayed for

MODE! key.... Parameters are not revised and the next item is displayed.

CALI key........Parameters are stored in the indicator and

Snd

I is displayed.

Press the jOperate/Standbvl key to turn the display off.

Press and hold the ¡NET/GROSSl key and press the |CAU key

....... .............All parameters are not changed

and the calibration mode is.

finished. Press the jOperate/Standby

key to turn the display off after

Turn the display on

for 10 miniuies

displaying

CBnCEL

CAL

m

,n

Zero

Calibration^

^

Span Calibration.

5. Calibration

AD-4329 Weighing Indicator

Page 13