5.3.1,

Setting Division and Range

Consider the foilowing rules to design the weighing range.

Rule 1 Select the division and range of each weighing range so as to fit the following inequality.

The first range < the second range < the third range

The division of next weighing range is automaticaliy set bigger than division of

lower weighing range. And the division can change.

Rule 2 When setting the dual range, the upper limit value of the second range becomes

the maximum capacity.

When setting the triple weighing range, the upper limit value of the third range

becomes the maximum capacity.

Rules Select a resolution smaller than 10000. The resolution is a value divided the

maximum capacity by the minimum division of the first range.

Mark to indicate the weighing range while this setting

w'

mark

First range

Second range

Third range

Net

ON

Gross

OFF

Procedure

Step 1 Press the |CAL| key to enter the calibration mode.

Then

CHL

i n

is displayed for two seconds.

Normal weighing mode

item Is displayed.

CAM key....... Parameters are stored in the indicator and

is displayed. Press the

End

key to turn the display off.

CAL

[ f ! L

i n

(

To Next page. A)

Press and hold the I

neitgross

I key and press the |CAl| key

All parameters are not changed and the

calibration mode is finished. Press the

Qperate/Standbvl key to turn the display off after

displaying I EdnEEl

AD'4329 Weighing indicator