5.2.7. Span Calibration

Step10 Check the | Fxxxx

I

display. The xxxx is a capacity.

Select a span caliiDration method to adjust the capacity.

Weighing a mass except the

maximum capacity

The method to weigh a mass except
the maximum capacity.

To step 11

Weighing maximum capacity

The method to weigh a mass

equivalent to the maximum capacity.

To step 14

Digital input

The numerical input to enter a load

ceil output voltage.

To step 16

Weighing a Mass except the Maximum Capacity

Step11 Set a mass value using the foilwing keys,

key.....Key to select a figure.

>

key Key to select a number.

Step12 Place a mass equivalent to displayed value on the

weighing unit.

StepIS Press the I ^ j key after the MD mark turns off.

Proceed to step 17.

MOPg key........The key not to change parameter

and go back to step 4.

CAll key...........The key not to change parameter

and proceed to step 17.

Caution Do not press the

J

key white the MD mark Is

tit (detecting motion). Arrange the condition
using the FOQ filter function.

Weighing Capacity Mass

Step14 Place a mass equivalent to the maximum capacity on

weighing unit.

StepIS

Press the ] key after the MD mark turns off.

Proceed to step 17.

MODE! key...... The key not to change parameter

and go back to step 4.

CALI key..........The key not to change parameter

and proceed to step 17.

key while the MD mark is

Caution Do not press the

lit (detecting motion). Arrange the condition
using the FOO filter function.

span calibration

Maximum capacity

Place the displayed

mass on the

weighing unit.

^ Unlit MD mark

5. Calibration

Page 16