9. Comparison

This function has the "upper / lower comparison" and the "setpoint comparison". They

compare the weighing data with preset parameters and can output the result of the

comparison to the reiay-outputs of OP-02.

□ Set the F-function the/^*¿'0 and /to use these comparisons.

Caution □ The result can only be output using relay-outputs, and is not displayed.

□ The setpoint comparison can be used for normal batching, but can not

be used for loss-in-weight.

9.1.

Comparator Output Relay (OP-02)

□ This relay option outputs the result of comparison controled by the F-function

f~^0

(comparator function) and / (comparator mode).

Specifications

250V AC or 30V DC

ЗА (resistive load)

100,000 times (resistive load)

М3

Maximum voltage

Maximum current

Contact life

Adaptable terminal type

Circuit and pin layout

Relay

7 7 7

Inside of OP-02

6.2

6.2

3.2

3.2

1 2

3 4

HI OK LO Common

Adaptable

Compression

Terminal

9.2.

Upper/Lower Limit Comparison

This function compares the weighing data with the upper limit value and lower limit

value, and outputs the result to the three relays of HI, OK and LO. Use this comparison

when judging whether a weight is proper.

Set to the F-function

f~EQ

I

to use this upper/ lower limit comparison.

Select a parameter of the F-function

I

for the comphson condition.

Specify the upper and lower limit values.

When inputting the upper value and lower limit value, it is not necessary to input the F-

function

f ~ BQ

and

f 1

again unless changing the F-function,

9. Comparison

Page 36

AD-4329 Weighing indicator