Zero calibration – A&D Weighing Indicator AD-4329 User Manual

Page 17

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

5.2.6.

Zero Calibration

Step 7 Check the

CfiL 0

display.

Select a zero calibration method to adjust the zero point

Weighing input

(Normal way)

The adjustment method

with nothing on the

weighing unit.

To step 8

Digital input

The numerical input to

enter a load cell output

voltage.

To step 9

Weighing Input

Step 8 Place nothing on the weighing unit. Press the

key

after the MD mark has turned off. Proceed to steplO.

MOPg key....... The key not to change the zero point

data and proceed to step 10.

Caution Do not press the [ *-■ key while the MD mark is lit

(detecting motion). Arrange the condition using

the FOO filter function.

Digital Input

Step 9 Pressing the @ key, a stored input voltage parameter

of the zero point is displayed in the unit of mV/V. Adjust

the input voltage using the following keys.

key.....The key to select a figure,

key.....The key to select a number.

key.....The key to store the zero point parameter

and proceed to step 10.

MODE] key..The key not to change parameter and

proceed to step 10.

Zero calibration

To Span Calibration

AD-4329 Weighing Indicator

Page 15

5. Calibration