background image

5.5. Gravity Compensation Function

□ If the scale is used at the caiibraiion location, it is not necessary to perform this function.

□ it may cause a weighing error, if there is a difference of gravity acceleration

between the installed location and calibration location. This function specifies

these gravity accelerations and corrects span error.

Caution □ The decimal point is not displayed in the function. Ex,

313B =

9.798 m/s^

Step 1 Press the ICALl key in normal weighing mode. The

Normai weighing mode

I

C3L in

is displayed for two seconds. Press the 1M+

key to enter the gravity compensation function.

If you want to cancel the current procedure, press

and hold the |NEi/GR0SS| key and press the |CAL| key. Then,

all parameters are not changed and the calibration

mode is finished. Press the lOperate/Standb^ key to turn the

display off after displaying

CRnCEL

Step 2 The parameter

0

1

xxxx is displayed. Enter the

>

gravity acceleration of the calibration location. The

parameter xxxx is the gravity acceleration.

key.......... The key to select a figure.

key...........The key to select a number.

key.......... The key to store the new gravity

acceleration and proceed to step 3.

A

mode

! key......... The key not to change the value and

proceed to step 3.

Step 3 The parameter

OB

xxxx is displayed. Enter the

gravity acceleration of the installed location. The

parameter xxxx Is the gravity acceleration.

>n key............ The key to select a figure.

7T| key............ The key to select a number.

key...........The key to store the new gravity

acceleration and proceed to step 4.

mode

! key......... The key not to change the value and

proceed to step 4.

Step 4 Now [ ¿W xxxx I is displayed. Press the ICAl] key to

store the parameters. The I

End

I is displayed.

Proceed to steps.

Step 5 Press the loperate/Siandby key to turn display off.

AD-4329 Weighing Indicator

Page 23

2 seconds

Calibrated location

Installed location

CAL

[Ri in

d”0P

M4*

6 I 3000

o

ZERO

>

-V|V

TARE

A

0 t 3138

Q

PRINT

08 3000

ZERO

>

TARE

A

08 3800

Q

PRINT

0 1

3198

CAL

End

I/O

0PR/ST8

Turn display off

5.5. Gravity CompensaBon Function