9,2,3. Setting Upper/Lower Limit Values

Step 1 Press the loperaie/Standbyl key to

turn

the display off.

Step 2

Press and hold the iPRINn key and press the

Qperate/Standbvl key to display J H j | for two seconds.

Step 3 Set the upper limit value (HI) using the following keys.

r> 1 key The key to select a figure.

A key.........The key to select a number.

+/-1 key.....The key to select plus or minus.

key..... The key to store the parameter and

proceed to step 4.

I

mod

I key ..The key not to change the parameter

and proceed to step 4.

S^6p 4

La

is displayed for two seconds.

I/(!)

m m

Display ofl

Q

PRINT I-

Press and hold

Press

1/Ф

0PR/ST8

2seconds

И

m/G

О

moss

Ш0

ТАЙН

>

A

Upper limit value

5000

Q

PRIfiT

2seconds I о

Uei/B/G

ZEROTARE

-f/*“

>

A

иш

Lower limit value.

Set the lower limit value (LO) using the same keys as step 3.

Step 5 Turn the dsiplay off using the ^pemte/Standbvl key.'

Q

PRINT

OPRiSTB

Display ofi

9.2.4, Example

Result

F-function:

f-PO

F'Pl

Upper iinit value:

5000

Lower Unit value:

1000

Condition Formula

Range

HI (Over)

OK (GO)

LO (Under)

5000 < Display value

5001

ON

OFF

1000

S

Display value ^ 5000

5000

1000

ON

5 < Display value < 1 GOO

999

6

ON

Display value ^5

5

OFF

9.2. Upper/Lower Limit Comparison