Specifications – A&D Weighing Indicator AD-4329 User Manual

Page 54

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

13. Specifications

Analog Input and A/D Conversion

input sensitivity

Up to 0,2 M-V/division

Input signal range

-1 mV-15 mV

Load cell excitation voltage

5V DC +5%, 120 mA with sense voltage input

Load cell drive capacity

Maximum 8 x 350

Q

load cells

Temperature coefficient

Zero

±{0.2 (iV + 0.0008 % of zero adjustment voItage)/C (typ.)

Span

±0.0008%/t: of reading (typ.)

Non-Linearity

0.01 % of full scale

Maximum input noise

Less than 0.4 ^tVp-p

input impedance

10 MQ or more

A/D conversion method

Integrating dual slope type

A/D resolution count

40000 counts

A/D conversion rate and

display update rate

Approximately 10 times/s

Maximum display resolution

10000 (permissible 20000)

Digital Section

Measurement display

7 segment* vacuum fluorescent display tube

Character color

Cobalt-blue

Character height

13 mm

State indicator

Symbol color

Cobalt-blue

Symbol

Minus sign, Zero point, Motion detect, Gross, Net,
Preset tare, Storing accumulation data

Unit

Character color

Green

Unit

Kg, t, lb

Interface

External Control Input

7 contact input terminals selected functions

Standard serial data output

Exclusive output for A&D peripherals

RS-232C interface

Serialinterface for communication

Relay output

3 point

Capacity: 250V AC, 30V DC. maximum current 3A (resistive load)
Comparison mode selection
HI, GO, LO output for upper/lower limit comparison

Zero band, preliminary, free fall, final for setpoint comparison

13. Specifications

Page 50

AD-4329 Weighing Indicator

Advertising