8.1.2.

Display and Operation

Action of Accumulating Data

□ When accumuiating data, the display blinks once.

If storing accumulation data, the M+ mark Is displayed.

Caution □ This function can not accumulate different unit data. Specify a unit before use.

Display of Accumulation Data

□ When setting to if - I / I (Effective accumulation function) and pressing the ¡MODE

key, the I

totRL I

is displayed and the total data is displayed with the M+ mark

blinking . If the I

mope

I key is pressed again, the weighing data Is displayed.

The total data can be output. Refer to "Output of Accumulation Data"

if "Ib/kg" of

[f-D I 3

Is selected, the display of "ib", "kg" and "total" is exchanged

with the IMOPEI key in order.

Undoing the Accumulation Data

□ The last weighing data can be deduced from the accumulation data unless new

data is accumulated.

and accumulation data.

Step 1

Step 2

Press the ¡MOPEj key to display

hatRL

Press and hold the g key more than 3 seconds. The display blinks once and the

last accumulation data before accumlating last weighing data is displayed.

Caution □ External input can not be used.

Clearing the Accumulation Data

Step 1 Press the I

mode

I key to display I totBl | and accumulation data.

Step 2 Press and hold the |ZERO| key more than 3 seconds. The display blinks once and

the accumuiation data is cleared.

Caution □ External input can not be used.

Output of Accumulation Data

□ Accumulation data can be output to the standard serial output and RS-232C

interface using the manual print mode.

Step 1 Press the |

mode

| key to display I frolrfff. I and accumulation data.

Step 2 Pressing the IPRIMH key, the count and total of accumulation data are output in order.

]

ySpace(20h)

Example: Count

f

4

. ojo 0 0 0|0I

\

.............. ^

Header

Data

Terminator

Total fTToiTiA|L|,kl0|0|0|0i0|0|likig|CR[h.|

Header

Data

Unit Terminator