C 150 (E2) 12/2012

page 15/90

The CO

2

incubators were produced in accordance with VDE regulations and were routinely tested in ac-

cordance to VDE 0411-1 (IEC 61010-1).

CAUTION

The glass door and the inner chamber will become hot during operation.

Danger of burning.

Do NOT touch the glass door, the inner surfaces or the charging material during opera-

tion.

WARNING

Stability hazard.

Danger of injury.

Damage to the unit and the charging material.

Housing cover breakaway.

Do NOT climb on the lower housing cover.

Do NOT load the lower housing cover with heavy objects while the unit door is open.

1.6

Precautions when working with CO

2

gas

Carbon dioxide (CO

2

) in high concentrations is hazardous to health. It is colorless and almost odorless

and therefore practically imperceptible. Vent out any CO

2

gas that may escape via good room ventilation

or a suitable connection to an exhaust system. We recommend installing a CO

2

warning system.

WARNING

High concentration of CO

2

(> 4 Vol.-%).

Risk of death by suffocation.

Danger of poisoning.

Do NOT set up units in non-ventilated recesses

 Ensure technical ventilation measures
 Observe the relevant regulations for handling CO

2

.