C 150 (E2) 12/2012

page 64/90

In case of eye contact, the disinfectant spray may cause eye damage due to chemical
burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-
fectant spray.

Recommended precautions: To protect the eyes use sealed protective goggles.

VORSICHT

Eye contact.

Eye damage due to chemical burns.

Do NOT empty into drains.

 Wear protective goggles.

Following frequent use of the disinfectant spray and prior to hot-air sterilization, remove any
agent remainder by using the neutral cleaning agent and then a moistened towel to avoid for-
mation of permanent residues.

After using the disinfectant spray, allow the incubator to dry thoroughly, and aerate it suffi-
ciently.

15.3 Disinfection of the CO

2

sensor

To ensure complete disinfection and proper function of the sensor, BINDER recommends a wipe disinfec-
tion of the sensor head with pure alcohol or non-corrosive alcohol-based surface disinfectants. The disin-
fectant must be non-corrosive and free of chlorine or any acid. We recommend using the disinfectant Art.
No. 1002-0022. Avoid strong shocks when handling the CO

2

sensor.

CAUTION

Excess temperature.

Immersion of sensor into liquids.

Shocks of the CO

2

sensor.

Damage to the CO

2

sensor.

Do NOT immerse the CO

2

sensor into liquids.

Do NOT expose the CO

2

sensor to autoclaving.

Do NOT expose the CO

2

sensor to hot-air sterilization.

 Avoid strong shocks of the CO

2

sensor (by putting it down hard, or dropping).

We recommend regular disinfection of the CO

2

sensor.

CAUTION

Connecting or removing the CO

2

sensor during operation.

Damage to the CO

2

sensor.

 Connect or remove the CO

2

sensor only with the unit turned off.