19.4 C 150 technical data

Exterior dimensions

Width

mm / inch

680 / 26.77

Height (incl. feet)

mm / inch

819/ 32.24

Depth

mm / inch

815 / 32.09

Depth plus door handle, I-triangle

mm / inch

54 / 2.13

Depth plus power supply connection and gas connec-
tion

mm / inch

60 / 2.36

Wall clearance rear

mm / inch

100 / 3.94

Wall clearance side

mm / inch

50 / 1.97

Number of doors

1

Number of inner glass doors

1

Interior dimensions

Width

mm / inch

500 / 19.69

Height

mm / inch

600 / 23.62

Depth

mm / inch

500 / 19.69

Interior volume

l

150 / 5.4

Number of shelves, series / max.

3/6

Size of shelf, width x depth

mm x mm

inch x inch

491 x 442

19.33 x 17.40

Weight (empty)

kg / lbs

95 / 209.5

Temperature data

Temperature range, 7 °C / 12.6 °F above ambient, up
to

°C / °F

50 / 122

Temperature fluctuation

± K

0.1

Temperature uniformity (variation) at 37 °C/ 98.6 °F

± K

0.4

Recovery time 1) after door open for 30 sec
at 37 °C/ 98.6°F

minutes

5

Humidity data

Humidity range

% r.H.

95 ± 2

CO

2

data

CO

2

range

vol.-% CO

2

0 to 20

Setting accuracy

vol.-% CO

2

0.1

Recovery time 1) after door open for 30 sec
at 5 vol.-% CO

2

minutes

7

CO

2

measurement

Drift-free CO

2

infrared absorption measurement

system

Connection hose nozzle DN6 for CO

2

Connection to the unit for hose with internal diameter

mm / inch

6 / 0.24

Electrical Data

IP system of protection acc. to EN 60529

IP

20

Nominal voltage (±10%) 50/60 Hz

V

230 1N~

Nominal power

kW

1.40

Energy consumption 2) at 37 °C/ 98.6°F

Wh/h

110

Power plug

Shock-proof plug

Installation category acc. to IEC 61010-1

II

Pollution degree acc. to IEC 61010-1

2

Unit fuse

5x20mm / semi time-lag / 10 A