BINDER C 150 User Manual

Page 37

Advertising
background image

C 150 (E2) 12/2012

page 37/90

Operating functions in operating mode USER

Display Setting range

Operating function

0.0 °C to 50.0 °C

Temperature set point
Must be at least 7 °C / 12.6 °F above the ambient temperature in order to
ensure a stable temperature inside the C 150.

0.0 vol.-% to 20.0
vol.-%

CO

2

set point

Common value: 5.0 vol.-% (factory setting), or depending on the NaHCO

3

concentration of the culture medium (Figure 27 page 55).

0.0 km to 2.0 km

Altitude of the installation site above sea level
For adjusting the displayed CO

2

concentration depending on the air pres-

sure

0.1 K to 10.0 K

Temperature alarm threshold
The maximum permitted deviation from the temperature set point in K,
Deviations trigger a tolerance range alarm (chap. 10.5) on the main con-
troller. For technical reasons, the deviation should be at least 1.0K.

0 to 999 min

Temperature alarm-delay time
Period until the main controller triggers a tolerance range alarm after
closing the door. This prevents alarms constantly being triggered during
the unstable operating phase after closing the door.
This has no effect on the safety controller, which can still trigger a limit
alarm or offset alarm.

0.1 vol.-% to 5.0
vol.-%

CO

2

alarm threshold

The maximum permitted deviation from the CO

2

set point in vol.-%. Devi-

ations trigger a tolerance range alarm on the main controller. For tech-
nical reasons, the value should be at least 1.0 vol.-%.

0 to 999 min

CO

2

alarm - delay time

Period until the main controller triggers a tolerance range alarm after
closing the door. This prevents alarms constantly being triggered during
the unstable operating phase after closing the door.

0.1 K to 10.0 K

Setting the offset value for door heating

Offset to the temperature set-point (inner chamber), standard setting 1.5
K. You can use this value to increase (setting 1.0 K) or decrease (setting
2.0 K) inner chamber humidity.

Advertising