C 150 (E2) 12/2012

page 35/90

Operating functions that can be set in operating mode HAND

Display Setting range

Operating function

0.0 °C to 50.0 °C

Temperature set point
Must be at least 7 °C / 12.6°F above the ambient temperature in order to
ensure a stable temperature inside the C 150.

0.0 vol.-% to 20.0
vol.-%

CO

2

set point

Common value: 5.0 vol.-% (factory setting), or depending on the NaHCO

3

concentration of the culture medium (Figure 27 page 55).

0.0 °C to 60.0 °C

Safety co

ntroller set point when set point type is “Limit” (Lit)

Limit value, i.e., maximum permitted absolute temperature value. When
exceeded, the safety controller triggers an alarm. For technical reasons,
the limit value should exceed the temperature set point by at least 2 °C.
Typical value: 39.0 °C / 102.2°F.

Or

0.1 K to 10 K

Safety controller set point when set point type is “Offset” (OFS)
Offset value, i.e. maximum over temperature above the selected set point
When exceeded, the safety controller triggers an alarm. For technical
reasons, the offset value should be at least 2 °C (factory setting).

0 to 999

Max. 3-digit numeric password for locking/unlocking the operating func-
tion setting in operating mode LOCK (chap. 7.4)

To permit temporary unlocking: select any password

0.

To permit permanent unlocking: set password to “0”.

The operating function “At.S” in operating mode USER (chap. 7.3) specifies the safety control-
ler set point

type “Limit” (Lit) or “Offset” (OFS).

With set point

type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit

value) is displayed in operating modes HAND and USER. You can set a limit value.

With set point

type “Offset” (OFS), operating function “Ot.S” (safety controller set point as

offset value) is displayed in operating modes HAND and USER. You can set an offset val-
ue.

With set point

type “Limit”, set the safety controller each time the set point for the temperature

is changed. Set the safety controller set point approx. 2 °C above the controller temperature
set point.

Factory setting: set point

type “Offset” with safety controller set point 2 °C.