BINDER C 150 User Manual

Page 38

Advertising
background image

C 150 (E2) 12/2012

page 38/90

Display Setting range

Operating function

0.0 °C to 60.0 °C

Safety controller set point when set point type is “Limit” (Lit)
Limit value, i.e., maximum permitted absolute temperature value. When
exceeded, the safety controller triggers an alarm. For technical reasons,
the limit value should exceed the temperature set point by at least 2 °C.
Typical value: 39.0 °C / 102.2°F.

Or

0.1 K to 10 K

Safety controller set point when set point type is “Offset” (OFS)
Offset value, i.e., maximum over temperature above the selected set
point. When exceeded, the safety controller triggers an alarm. For tech-
nical reasons, the offset value should be at least 2 °C (factory setting).

-----

Inhibit time. This operating function is not required to operate the C 150.
Its value should always be zero.

Set point type for safety controller: “Limit” (Lit)
In operating modes USER and HAND you can define the limit value
(“Al.S”).

Set point type for safety controller: “Offset” (OFS)
In operating modes USER and HAND you can define the offset value
(“Ot.S”).

0 to 999

Max. 3-digit numeric password for locking/unlocking the operating func-
tion setting in operating mode LOCK (chap. 7.4)

To permit temporary unlocking: select any password

≠ 0.

To permit permanent unlocking: set password to “0”.

0 to 999

Password setting for access to operating mode USER
Remember the modified password or you will no longer be able to access
operating mode USER.

1 to 254

Setting the unit address

Addressing serves for data transfer e.g. via the communication software
APT-

COM™ 3 DataControlSystem. The setting of "1" may not be

changed when the unit is fitted with a ethernet interface. When the unit
comes with a RS422-interface the settings of the Modbus address must
be done here.