Display

Setting range

Operating function

Firmware revision of main controller for service / maintenance pur-
poses.

Firmware revision of safety controller for service / maintenance pur-
poses.

Data record (year) for service / maintenance purposes.

Data record (month) for service / maintenance purposes.

Data record (day) for service / maintenance purposes.

Data set version for service / maintenance purposes.

OFF, or -199,9 °C
to +300 °C

Temperature adjustment serves to adjust the temperature sensor.

OFF, or
-10.0 vol.-% to
110.0 vol.-%

CO

2

final value adjustment serves to adjust the CO

2

sensor.

The operating function “At.S” specifies the safety controller set point type “Limit” (Lit) or “Off-
set” (OFS).

With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit
value) is displayed in operating modes HAND and USER. You can set a limit value .

With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as
offset value) is displayed in operating modes HAND and USER. You can set an offset val-
ue.