The settings of the safety controller are inactive during sterilization (chap. 15.4). They become
functional again following abortion of the sterilization and/or the restart of the unit at the main
power switch.

9.2.1 Setting the safety controller set point type

Select the safety controller set point type in operating mode USER (chap. 7.3). The operating function
At.S” specifies the safety controller set point type “Limit” (Lit) or “Offset” (OFS).

With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit value) is
displayed in operating modes HAND and USER. You can set a limit value.

With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as offset value)
is displayed in operating modes HAND and USER. You can set an offset value.

Normal Display

simultaneously for 3 seconds

Toggle to operating mode USER

Operating mode HAND

Operating mode USER

If operating mode HAND appears,
toggle to operating mode USER with
the arrow keys.

Password request

Password entry
(factory setting: “1”).