Caution – BINDER C 150 User Manual

Page 69

Advertising
background image

C 150 (E2) 12/2012

page 69/90

CAUTION

Glass door and inner chamber become hot during sterilization.

Danger of burning.

Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting steri-

lization.

Aborting sterilization after less than 6 hours

When aborting the sterilization prematurely, it may be that the cells/pathogens inside the unit have not all
been killed. You should repeat the sterilization.

CAUTION

Interruption of temperature reaction time.

Ineffective sterilization.
 Repeat the sterilization.

CAUTION

Glass door and inner chamber become hot during sterilization.

Danger of burning.

Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting steri-

lization.

When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in

the CO

2

sensor.

The CO

2

sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.

Turn on the unit (chap. 5).

Three events which can abort the automatic sterilization cycle prematurely:

A manual abortion (chap. 15.5.1)

Opening the outer door (chap. 15.5.2)

Turning off the power to the C 150 or a brief power failure (chap. 15.5.3)

15.5.1 Aborting the hot-air sterilization manually

You can abort the sterilization cycle manually.

Press the controller button “180 °C” for three seconds during the sterilization cycle.

The lower display will show the operating function “StE”, and the value “ON” will appear in the upper
display.

Press “▲”.

The upper display toggles from “ON” to “OFF“.

Press “MODE” to end the sterilization cycle.

Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F.