Warning, 2 base on castors (option), 1 analog outputs for temperature and co – BINDER C 150 User Manual

Page 58: Option)

Advertising
background image

C 150 (E2) 12/2012

page 58/90

WARNING

High concentration of CO

2

(> 4 Vol.-%).

Risk of death by suffocation.

Danger of poisoning.
 Tightly close each access port with two plugs during operation.

13.2 Base on castors (option)

In order to obtain easy access to the incubator and to avoid contamination of the incubator caused by soil
pollution, BINDER recommends using the base on castors.

The mounting instructions 7001-0147 delivered with the base on castors describe its
installation (Art. No. 9051-0024).

13.3 Stacking adapter for direct thermal decoupled stacking (option)

We recommend not stacking CO

2

incubators directly on top of one another in order to avoid transmission

of shocks and vibrations from one unit to the other. This could happen e.g. while opening or closing the
door, cleaning, loading and unloading the unit. BINDER offers a stacking adapter for direct thermal de-
coupled stacking of two incubators.

The stacking adapter ensures the exact maintenance of the set incubation parameters also during sterili-
zation of the other unit in the same stacking stand (chap. 15.4).

The mounting instructions 7001-0145 delivered with the stacking adapter describe its
installation (Art. No. 9051-0025 for 2 C150, Art. No. 9051-0027 for C 150 on top of CB
150).

13.1 Analog outputs for temperature and CO

2

(option)

With this option, the CO

2

incubator is equipped with analog outputs 4-20 mA for temperature and CO

2

.

These outputs allow transmitting data to external data registration systems or devices.

The connection is realized as a DIN socket at the rear of the CO

2

incubator as follows:

ANALOG OUTPUT 4-20 mA DC

PIN 1: Temperature

PIN 2: Temperature +'
PIN 3: CO

2

PIN 4: CO

2

+

CO

2

range:

0 vol.-% up to 20 vol.-%

Temperature range: 0 °C / 32°F up to +200 °C / 392°F

A suitable DIN plug is enclosed.

Figure 31: Pin allocation of DIN socket for option analog outputs

Advertising