P. 15.1, Danger, Caution – BINDER C 150 User Manual

Page 62: 1 cleaning

Advertising
background image

C 150 (E2) 12/2012

page 62/90

15. Cleaning, decontamination / disinfection, and sterilization

DANGER

Electrical hazard.

Danger of death.

Do NOT spill water or cleaning agents over the inner and outer surfaces.

 Before cleaning, turn off the unit at the main power switch (2) and dis-

connect the power plug.

 Completely dry the appliance before turning it on again.

15.1 Cleaning

Disconnect the CO

2

incubator from the power supply before cleaning. Disconnect the power plug.

Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:

Exterior surfaces,
instrument panel:

Standard commercial cleaning detergents free from acid or halides.

Alcohol-based solutions.

We recommend using the neutral cleaning agent Art. No. 1002-0016.

Inner chamber,
shelves,
water pan:

Standard commercial cleaning detergents free from acid or halides.

Copper sulphate solutions or alcohol-based solutions.

We recommend using the neutral cleaning agent Art. No. 1002-0016.

CO

2

sensor

Alcohol-based solutions

Do not immerse the CO

2

sensor into the solution.

Disinfection with alcohol or an alcohol-based surface disinfectant without corro-
sive effect, free from acid or halides. We recommend using the disinfectant spray
Art. No. 1002-0022.

Silicone door gasket: Alcohol-based solutions or neutral cleaning agent Art. No. 1002-0016.

Zinc coated hinge
parts, rear unit wall

Standard commercial cleaning detergents free from acid or halides. Do NOT use
a neutral cleaning agent on zinc coated surfaces.

We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
cleaning.

Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.

CAUTION

Danger of corrosion.

Damage to the unit.

Do NOT use acidic or chlorine cleaning detergents.

Do NOT use the neutral cleaning agent on other kind of surfaces e.g., the zinc coated

hinge parts or the rear unit wall.

For surface protection, perform cleaning as quickly as possible.

After cleaning completely, remove any cleaning agents from the surfaces with a moistened
towel. Let the unit dry.

Soapsuds may contain chlorides and must therefore NOT be used for cleaning.