Caution, Danger – BINDER CB 60 User Manual

Page 26

Advertising
background image

CB (E6) 09/2013

Page 26/144

CAUTION

Danger of overheating.
Damage to the unit.

Do NOT set up units in non-ventilated recesses.

Ensure sufficient ventilation for dispersal of the heat.

Do not install or operate the CO

2

incubator CB in potentially explosive areas.

DANGER

Explosion hazard.
Danger of death.

Do NOT operate the unit in potentially explosive areas.

KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.


Ambient conditions

Permissible ambient temperature range for operation: +18 °C / 64.4 °F to +30 °C / 86 °F

At elevated ambient temperature values, fluctuations in temperature can occur.

Ideal ambient temperature: at least 7 °C / 12.6 °F below the intended working temperature. E.g.,

working temperature 37 °C / 98.6 °F = ambient temperature 30 °C / 86 °F and lower. In the event of

working temperatures of less than 7 °C / 12.6 °F above the ambient temperature, the setpoint can be

exceeded.

The ambient temperature should not be substantially higher than the indicated ambient

temperature of 22 ±3 °C / 71.6 ±5.4 °F to which the specified technical data relates. For other

ambient conditions, deviations from the indicated data are possible.

Avoid direct solar radiation on the unit.

Permissible ambient humidity: 70 % r.H. max., non-condensing.

Installation height: max. 2000 m / 6561.7 ft above sea level.

Wall distances: rear 100 mm / 3.94 in, sides 50 mm / 1.97 in.