15.5 Analog outputs for temperature and CO

2

(option)

With this option, the CO

2

incubator is equipped with analog outputs 4-20 mA for temperature and CO

2

.

These outputs allow transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (3) on the unit rear. A suitable DIN plug is enclosed.

ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature +

PIN 2: Temperature -

PIN 3: CO

2

+

PIN 4: CO

2

-

CO

2

range:

0 vol.-% up to 20 vol.-%

Temperature range: 0 °C / 32°F up to +200 °C / 392°F

Figure 21: Pin configuration of the DIN socket (3) on the unit rear

15.6 Access port for extra-low voltage (option)

The access port (B) (8-pin) for extra-low voltage (ELV) consists of a LEMO socket (which can be covered)

and a LEMO connector. It is suitable to connect two electrical devices, one inside and one outside the

incubator. You can use it e.g. for devices with a monitoring and control part remaining outside the

incubator, whereas its mechanical component like roller or shaker systems is located inside the unit.

Figure 22: Position of the access port for extra-low voltage


Maximum power rating 24V AC/DC 2Amp.

WARNING

Overload of contacts.
Damage to contacts and connection socket.
Electrical hazard.
Danger of death.

Do NOT exceed the maximum load of 24V AC/DC 2Amp.

Do NOT connect any devices with a nominal current > 2 Amp.

Do NOT connect any devices with a nominal voltage > 24 V DC.