BINDER CB 60 User Manual

Page 75

Advertising
background image

CB (E6) 09/2013

Page 75/144


Icon

Alarm message

Signification

Humidity range

After the humidity was situated within the defined tolerance range (chap.

11.5.6) it deviates longer than 10 min. from the tolerance range

or the

humidity doesn’t reach the tolerance range within 3 hours from turning on

the unit or closing the door

Door open

Door is open

for more than the set alarm delay time (chap. 11.5.5, factory

setting: 1 minute). Close the door.

Low pressure CO2 Low CO

2

outlet pressure (< 0.3 bar)

Low pressure CO2 Low O

2

outlet pressure (< 0.3 bar) (unit with O

2

control)

Low pressure N2

Low N

2

outlet pressure (< 0.3 bar) (unit with O

2

control)

CO2 sensor

defective

CO

2

sensor defective. Contact BINDER Service.

O2 sensor

defective

O

2

sensor defective. Contact BINDER Service.

Inform service -

adjust CO2 sensor

CO

2

sensor adjustment required. Contact BINDER Service.

You can activate / deactivate the buzzer in the “Alarms” submenu (chap. 11.5.3).
With an activated buzzer there is an

audible alert with an alarm. You can reset it in the “Alarms”

submenu for alarm acknowledgement pressing the “Reset” button (chap. 11.4). The alarm symbol will

only disappear when the cause of the alarm has been remedied.
The

zero-voltage relay alarm contact (chap. 11.6) is activated in case of the alarm messages “door

open” and deviations of temperature and gas concentrations) as well as in case of a power failure and

turning off the incubator at its main power switch.
For appropriate actions in the event of an alarm, please refer to chap. 21 “Troubleshooting”.

Advertising
This manual is related to the following products: