CAUTION

Excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the unit.

The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.

Before connecting, check the outlet pressure on the pressure reducer of the

cylinder.

Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure.

Observe the correct outlet pressure also when replacing the gas cylinders.

Establishing the connection to the incubator

Connect the supplied gas hose (internal diameter 6 mm / 0.24 inches) to the pressure reducer of the

gas cylinders or central gas supply and secure the connection with the supplied hose clamp.
Connect the pre-assembled hose nozzle of the gas hose to the quick acting closure socket (6) DN 6

on the unit rear, as described in chap. 4.3.4.

Leak test

After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or diluted

soap solution).

The recovery times of the gas concentrations inside the chamber after opening the door are

indicated in the technical data (chap. 22.4) and refer to a connection pressure of 2.0 bar / 29

psi. Decreasing supply pressure will result in longer recovery times.

Conversion table for gas inlet pressures, bar – psi, see chap. 22.8.

4.3.4 Connecting the gas hose to the unit rear (for CO

2

, O

2

, N

2

)

The procedure of connecting the gas hose to the unit rear is the same for any gas connection. All quick

acting closure sockets (CO

2

, and N

2

and O

2

on unit with O

2

control) are degreased and supplied with a

FKM gasket.

Note for units with O

2

control: The quick acting closure socket and the hose nozzle for the O

2

connection (unit with O

2

control) must be degreased.

WARNING

Fire and explosion hazard through contact of fat with O

2

.

Risk of burns and other injuries.

All connection parts for the O

2

connection must be degreased.

The gas hose, which will be used to establish the connection to a gas cylinder, is already attached to the

hose nozzle and secured by a hose clamp. Plug the hose nozzle into the corresponding quick acting

closure socket (a) located at the rear of the unit. This quick acting closure socket is closed by a rubber

cover (b).

Only use the supplied hose nozzle to connect to the quick acting closure socket.
Otherwise, the quick acting closure socket may leak, and/or it may become impossible to

connect the original hose nozzle. In this case, please contact BINDER Service.