CB (E6) 09/2013

Page 53/144


Setting the CO

2

concentration

To enter the CO

2

setpoint, go to

Quick menu > CO2

setpoint

Fixed value

08.03.2013 05:05:06

Entry menu “CO2 setpoint”.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 0 vol.-% up to 20 vol.-%.
Press the “Ok” button to confirm.

..\ CO2 setpoint

5

.0

[Vol.-%]

5

End Ok 0 1 2 3 4

6 7 8 9 Ins Del

Close

Ok

Home

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.
Go back to the initial view with “Home” or enter the O

2

concentration (unit with O

2

control).


Setting the O

2

concentration (unit with O

2

control)

To enter the O

2

setpoint, go to

Quick menu > O2

setpoint

Fixed value

08.03.2013 05:05:06

Entry menu “O2 setpoint”.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 0.2 vol.-% up to 95 vol.-%.
Press the “Ok” button to confirm.

..\ O2 setpoint

5

.0

[Vol.-%]

5

End Ok 0 1 2 3 4

6 7 8 9 Ins Del

Close

Ok

Home

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.
Go back to the initial view with “Home”.