1 required gas supply of the unit with o2 control, 3 deactivating the o, Control and o – BINDER CB 60 User Manual

Page 45: Pressure alarms (unit with o, Control)

Advertising
background image

CB (E6) 09/2013

Page 45/144

6.3 Deactivating the O

2

control and O

2

/N

2

pressure alarms (unit with O

2

control)

6.3.1 Required gas supply of the unit with O

2

control

In the unit with O

2

control, cell growth can be additionally influenced by a variable oxygen content.

Depending on the required range of O2 concentration, the type of gas which is required may vary:

Hyperoxic range (> 22 vol.-% O

2

)

As compared to the ambient air (O

2

concentration of 20.9 vol.-%) only the O

2

concentration needs to be

increased, no nitrogen is required for the O

2

control. Therefore, it is not necessary to connect a N

2

cylinder. You can deactivate the N

2

pressure alarm in the menu “

Menu > Controller mode > O2 control

(chap. 6.3.2).

Hypoxic range (< 20 Vol.-% O

2

)

As compared to the ambient air (O

2

concentration of 20.9 vol.-%) only the N

2

concentration needs to

be increased, no oxygen is required for the O

2

control. Therefore, it is not necessary to connect an O

2

cylinder. You can deactivate the O

2

pressure alarm in the menu “

Menu > Controller mode > O2

control

” (chap. 6.3.2).

Hyperoxic and hypoxic range or range from 20 vol.-% O

2

up to 22 vol.-% O

2

N

2

and O

2

are required for O

2

control. It is necessary to connect all gas supplies.

Operation without O

2

control

You can deactivate the O

2

control in the menu “

Menu > Controller mode > O2 control

” (chap. 6.3.2).

With this setting, no alarm messages for O

2

and N

2

(pressure or concentration) will be issued.

Since neither nitrogen nor oxygen are required, it is not necessary to connect an O

2

or N

2

cylinder. No

O

2

and N

2

pressure alarms will be issued with deactivated O

2

control.

The zero-voltage relay alarm contact (chap. 11.6) for messages of deviation of the O

2

concentration

will also not be triggered.


Overview

Hyperoxic range

(>22 vol.-% O

2

)

Hypoxic range

(<20 vol.-% O

2

)

Hyperoxic and

hypoxic ranges

No O

2

control

O

2

control (with O

2

concentration alarm)

active

active

active

off

Connection of O

2

cylinder

connected

not connected

connected

not connected

O

2

pressure alarm

active

off

active

off

Connection of N

2

cylinder

not connected

connected

connected

not connected

N

2

pressure alarm

off

active

active

off

Advertising
This manual is related to the following products: