BINDER CB 60 User Manual

Page 46

Advertising
background image

CB (E6) 09/2013

Page 46/144

6.3.2 Activating / deactivating the O

2

control and O

2

/ N

2

pressure alarms

To activate or deactivate the O

2

control and O

2

/ N

2

pressure alarms, go to

Menu > Controller mode > O2 control

Fixed value

08.03.2013 05:05:06

Submenu “O2 control”.
The current settings are displayed.
To change the settings, press the operating button.
The modified settings are displayed

“O2 control: On” = O

2

control is active.

“O2 control: Off” = O

2

control is deactivated.

“O2 pressure alarm: On” or “N2 pressure alarm: On”

= the corresponding pressure alarm is activated.
“O2 pressure alarm: Off” or “N2 pressure alarm: Off”

= the corresponding pressure alarm is deactivated.

..\ O2 control

O2 control: On

O2 pressure alarm: On

N2 pressure alarm: On

Close

Home

Go back to the initial view with “Home”.

Fixed value

08.03.2013 05:05:06

Initial view with deactivated O

2

control

(sample values).
Also the pressure alarms for O

2

and N

2

are

deactivated.

Setpoint

Actual value

Temperature [°C]

37.0 37.2

CO2 [Vol.-%]

5.0

5.0

Humidity [%rH]

85.5

O2 [Vol.-%]

20.7

User

Quick menu

Menu

To enter the O

2

setpoint, first activate the O

2

control again.


With active O

2

control, the pressure alarms for O

2

and N

2

can be deactivated individually.

Fixed value

08.03.2013 05:05:06

Initial view with activated O

2

control and deactivated

O

2

pressure alarm (sample values).

Setpoint

Actual value

Temperature [°C]

37.0 37.2

CO2 [Vol.-%]

5.0

5.0

Humidity [%rH]

85.5

O2 [Vol.-%]

25.0 25.0

User

Quick menu

Menu