3 alarm status – BINDER CB 60 User Manual

Page 76

Advertising
background image

CB (E6) 09/2013

Page 76/144

11.3 Alarm status

An alarm message can appear in 3 different states:

“Set”

Active alarm.

The corresponding alarm icon is displayed in the initial view The buzzer sounds (if activated).

The “Info” button in the initial view leads to the “Alarms” submenu for alarm acknowledgement.

Press the “Reset” button in the “Alarms” submenu for alarm acknowledgement to mute the buzzer and

confirm the alarm.


“Acknowledged”

Active alarm.

The alarm was acknowledged. The buzzer is off.

The alarm cause is still valid. Therefore, the alarm icon remains displayed in the initial view.

The alarm message figures in the list of active alarms.


“Cleared”

The alarm cause has disappeared.

The alarm icon is not displayed any longer

The alarm message has disappeared from the list of active alarms.

The alarm message remains in the “Event list” for information.