Table 1-1, Product matrix -2, Model 3081fg – Emerson 3081FG User Manual

Page 12: Instruction manual, Table 1-1. product matrix

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5
September 2002

1-2 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

Table 1-1. Product Matrix

3081FG

High Temperature Oxygen Flue Gas Analyzer

High Temperature Analyzer - Instruction Book

Code

Sensing Probe Length

1

20 in. (508 mm) probe, 1/4 in. tube fittings

2

26 in. (660 mm) probe, 1/4 in. tube fittings

3

34.625 in. (880 mm) probe, 1/4 in. tube fittings

Code

Probe Outer Tube Material - Maximum Operating Temperature

1

Alumina - 2912

°F (1600°C) maximum - 1.25 NPT mounting

2

Inconel 600 - 1832

°F (1000°C) maximum - 1.25 NPT mounting

Code

Mounting Adapter - Stack Side

0

No adapter plate required uses 1.25 NPT
("0" must also be chosen under "Mounting Adapter" below)

1

New flanged installation - Square weld plate with studs (matches "Mounting Adapter" below)

2

Model 450 mounting ("4" must also be chosen under "Mounting Adapter" below)

3

Competitor's Mount ("5" must also be chosen under "Mounting Adapter" below)

Code

Mounting Adapter - Probe Side

0

No adapter plate

1

ANSI 2 in. 150 lb flange to 1.25 NPT adapter
(6 in. dia. flange, 4.75 in. BC with 4 x 0.75 in. dia. holes)

2

DIN to 1.25 NPT adapter (184 mm flange, 145 mm BC with 4 x 18 mm dia. holes)

3

JIS to 1.25 NPT adapter (155 mm flange, 130 mm BC with 4 x 13 mm dia. holes)

4

Model 450 to 1.25 NPT adapter

5

Competitor's mounting flange

Code

Electronics & Housing - Intrinsically Safe, NEMA 4X, IP65

1

3081 Electronics (Hart-compatible) - CENELEC EEx ia IIC T5

2

3081 Electronics (Hart-compatible) - CSA pending

3

3081 Electronics (Hart-compatible) - FM Class I, Div. I, Groups B,C,D

Code

Housing Mounting

0

Surface or wall mounting

1

1/2 to 2 in. pipe mounting

Code

Communications

0

No remote control

1

Infrared Remote Control (IRC)
(LCD display through cover window)

Code

Calibration Accessories

1

No hardware

2

Calibration and reference air flowmeters and refer-
ence air pressure regulator

Code

Armored Cable Length

00

No cable

11

20 ft (6 m)

12

40 ft (12 m)

13

60 ft (18 m)

14

80 ft (24 m)

15

100 ft (30 m)

16

150 ft (45 m)

17

200 ft (61 m)

18

300 ft (91 m)

19

400 ft (122 m)

20

500 ft (152 m)

3081FG

2

1

0

0

1

1

1

2

11

Example

Advertising