Troubleshooting 6-5

Model 3081FG

%

26020026

Figure 6-7. Fault 3, Shorted Thermocouple

c. Fault 3, Shorted Thermocouple

The thermocouple connections are shorted.
The fault displays as shown in Figure 6-7.

1. Allow adequate time for the oxygen

probe to reach operating temperatures.
Probe temperatures below approxi-
mately 500

°C (932°F) may result in this

fault.

2. Refer to Figure 6-4. Using a multime-

ter, measure between terminals 8(-)
and 10(+).

3. If the multimeter reading, in voltage

mode, is between -0.5 and +0.5 mV,
the thermocouple is shorted.

4. If the thermocouple is shorted, replace

the oxygen probe per paragraph 5-2.

5. If the thermocouple is not shorted, then

replace the PC board stack assembly
per paragraph 5-1.

%

26020027

Figure 6-8. Fault 4, High Probe Temperature

d. Fault 4, High Probe Temperature

The probe’s temperature has exceeded the
maximum cell temperature setpoint. The
fault displays as shown in Figure 6-8.

1. If the probe temperature exceeds the

maximum cell temperature setpoint,
the 4-20 mA signal output will become
invalid and go to the default value.

2. Verify that the upper cell temperature

setpoint is configured as desired under
the PROGRAM MENU (see Section 3,
STARTUP AND OPERATION).