Program menu, Program menu -6, Figure 3-7. code -6 figure 3-8. display code -6 – Emerson 3081FG User Manual

Page 36: Model 3081fg, Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5
September 2002

3-6 Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

3-5

PROGRAM MENU

The PROGRAM MENU branch of the menu
tree allows you to program and edit some proc-
ess parameters, faults, outputs, and security
codes. To access this branch of the menu tree,
press the PROG key on the IRC when in the
PROCESS DISPLAY screen (Normal or
Faulted). If security is enabled, you must enter
the analyzer code to gain further access to the
screens in this branch. Each screen in this
branch is accessed sequentially using the NEXT
key. Refer to Figure 3-5 during the following
menu and screen descriptions.

NOTE

To edit a screen value, press ENTER to
access the data field. Use the left and
right arrow keys to move among the
digits in the data field. Note that the
editable position will be flashing. To
change the value of a digit, use the up
and down arrow keys to increase or
decrease the value. When finished ed-
iting, press ENTER to accept the value.
To go to the next screen in the menu,
press NEXT.

a. CODE

Refer to Figure 3-7. After pressing the
PROG key, this screen will display if secu-
rity is enabled (see paragraph 3-5m). Use
this screen to identify a specific analyzer in
a process to prevent accessing an adjacent
analyzer when using the IRC.

Press ENTER to begin editing. At this point,
you can either specify the analyzer by its
access code or view its code if it is un-
known.

1. To gain further access to the screens

in the PROGRAM MENU branch, en-
ter the correct three-digit analyzer ac-
cess code using the arrow keys and
press ENTER. If security is disabled,
this screen does not appear and the
system displays the FAULT VAL
screen.

2. If the analyzer access code is un-

known, enter 555 and press ENTER to
access the DISPLAY CODE screen. In
that screen, you will be able to view the
analyzer access code.

%

O CONCENTRATION

2

ANALYZER

ACCESS CODE

OR SECURITY

ACCESS CODE

26020009

Figure 3-7. CODE

b. DISPLAY

CODE

Refer to Figure 3-8. This screen is accessi-
ble from the CODE screen by entering 555
and pressing ENTER. The DISPLAY CODE
screen identifies the analyzer access code
so you can return to the CODE screen and
enter the code as described in paragraph
3-5a. To return to the CODE screen, press
NEXT.

%

O CONCENTRATION

2

ANALYZER

ACCESS CODE

26020047

Figure 3-8. DISPLAY CODE