Model 3081fg, Instruction manual – Emerson 3081FG User Manual

Page 48

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5
September 2002

4-2 HART/AMS

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

LOOP CONNECTORS

USE INTERFACE

00275 0013 ONLY

SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS

SERIAL PORT

MODEL 3081 TRANSMITTER

TERMINAL BLOCK

LOOP CONNECTORS

4-20 mA SIGNAL LINE

HART

COMMUNICATOR

RL

250

³

W

ANALOG OUTPUT DEVICE

LEAD SET

HART COMMUNICATOR

REAR PANEL

12

3

5

6

7

8

9

11

12

13

14

15

16

10

4

+

4-20 mA

26020040

Figure 4-1. Signal Line Connections, > 250 Ohms Lead Resistance

a. Method 1, For Load Resistance

≥≥≥≥ 250

Ohms

Refer to Figure 4-1 and the following in-
struction to connect the HART Communi-
cator to a signal line with 250 ohms or more
of load resistance.

Explosions can result in death or seri-
ous injury. Do not make connections
to the HART Communicator's serial
port, 4-20 mV signal line, or NiCad re-
charger jack in an explosive
atmosphere.

Using the supplied lead set, connect the
HART Communicator in parallel to the Two-
Wire In Situ Oxygen Analyzer. Use any wir-
ing termination points in the analog output
4-20 mA signal line.

b. Method 2, For Load Resistance < 250

Ohms

Refer to Figure 4-2 and the following steps
to connect the HART Communicator to a
signal line with less than 250 ohms load
resistance.

Explosions can result in death or seri-
ous injury. Do not make connections
to the HART Communicator's serial
port, 4-20 mA signal line, or NiCad re-
charger jack in an explosive
atmosphere.

1. At a convenient point, break the analog

output 4-20 mA signal line and install
the optional 250 ohm load resistor.

2. Plug the load resistor into the loop

connectors (located on the rear panel
of the HART Communicator).