1-6 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

DUCT

STACK

GASES

PRESSURE

REGULATOR

FLOWMETER

OPTIONAL

ADAPTER

PLATE

4-20 mA SIGNAL

26020038

MODEL 3081

TRANSMITTER

OXYGEN

PROBE

INSTRUMENT

AIR SUPPLY

(REFERENCE AIR)

Figure 1-3. Typical System Installation

A source of instrument air is required at the
oxygen probe for reference air use. Since
the Two-Wire In Situ Oxygen Analyzer is
equipped with an in-place calibration fea-
ture, provisions should be made for con-
necting calibration check gas tanks to the
oxygen probe during calibration.

If the calibration check gas bottles are to be
permanently connected, a check valve is
required next to the calibration fittings on
the integral electronics.

This check valve is to prevent breathing of
calibration check gas line and subsequent
flue gas condensation and corrosion. The

check valve is in addition to the stop valve
in the calibration check gas kit.

NOTE

The electronics of the Model 3081
Transmitter is rated NEMA 4X (IP65)
and is capable of operating at tem-
peratures up to 65°C (149°F).

Retain the packaging in which the
Rosemount Two-Wire In Situ Oxygen
Analyzer arrived from the factory in
case any components are to be
shipped to another site. This packag-
ing has been designed to protect the
product.