Model 3081fg, Instruction manual – Emerson 3081FG User Manual

Page 37

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-7

Model 3081FG

%

mA

O CONCENTRATION

2

INDICATOR

FAULT VALUE

26020010

Figure 3-9. FAULT VAL

c. FAULT VAL

Refer to Figure 3-9. Use this screen to set
the value that the 4-20 mA output will drive
to and display during a fault condition. Press
ENTER to begin editing. Use the arrow keys
to enter a fault value. The fault value can be
between 3.8 and 24 mA. Then, press EN-
TER to accept the value. Pressing NEXT
displays the UPPER RANGE VAL screen.

Refer to Section 6, TROUBLESHOOTING,
for the actual fault conditions.

d. UPPER RANGE VAL

Refer to Figure 3-10. Use this screen to set
the value of the upper range limit. This
value is the maximum limit of the O

2

con-

centration measurement and is used to
scale the 4-20 mA output. Press ENTER to
begin editing. Use the arrow keys to select
and change the value. The upper range
value can be between 0 and 25%. Then,
press ENTER to accept the value. Pressing
NEXT displays the CELL T HI screen.

e. CELL T HI

Refer to Figure 3-11. Use this screen to set
the value of the upper cell temperature fault
condition. This value is the maximum al-

lowed cell temperature before a fault condi-
tion is indicated. Press ENTER to begin
editing. Use the arrow keys to select and
change the value. The value must be be-
tween 550° and 1600°C. Press ENTER to
accept the value. Pressing NEXT displays
the RESET MAX CELL T screen.

%

%

O CONCENTRATION

2

UPPER

RANGE LIMIT

26020011

Figure 3-10. UPPER RANGE VAL

%

O CONCENTRATION

2

26020012

UPPER CELL

TEMPERATURE

FAULT VALUE

Figure 3-11. CELL T HI

3