Model 3081fg, Instruction manual – Emerson 3081FG User Manual

Page 44

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5
September 2002

3-14 Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

a. IN MANUAL?

Refer to Figure 3-27.

Failure to remove the analyzer from
automatic control loops prior to per-
forming this procedure may result in a
dangerous operating condition.

If the O

2

output value is used in any auto-

matic process control loops, the loop must
be placed in manual before beginning a
calibration check.

Once the analyzer is removed from any
automatic control loops, press ENTER to
edit the screen. Use the arrow keys to se-
lect Y (yes) and press ENTER to process
the selection and to display the ACCEPT
HIGH O

2

screen.

%

O CONCENTRATION

2

Y/N (FOR USER

RESPONSE)

26020054

Figure 3-27. IN MANUAL?

%

%

O CONCENTRATION

2

HIGH CALIBRATION

CHECK GAS O VALUE

2

26020055

Figure 3-28. ACCEPT HIGH O

2

b. ACCEPT HIGH O

2

Refer to Figure 3-28. After pressing ENTER
to begin the calibration check, the high cali-
bration check gas starts to flow. After wait-
ing approximately three minutes for the
displayed O

2

value to settle, press NEXT to

accept the high calibration check gas read-
ing and apply the low calibration check gas.
The next screen to display is the ACCEPT
LOW O

2

screen.

c. ACCEPT LOW O

2

Refer to Figure 3-29. Once the low calibra-
tion check gas is applied, wait approxi-
mately three minutes for the displayed O

2

value to settle. Once the value settles, press
NEXT to accept the reading and to display
the SLOPE screen.