Startup and operation, General, Power up – Emerson 3081FG User Manual

Page 31: Startup and operation -1, General -1, Power up -1, Model 3081fg, Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-1

Model 3081FG

SECTION 3

STARTUP AND OPERATION

Install all protective equipment covers
and safety ground leads before
equipment startup. Failure to install
covers and ground leads could result
in serious injury or death.

3-1 GENERAL

a. Verify Mechanical Installation

Ensure the Two-Wire In Situ Oxygen Ana-
lyzer is installed correctly. See paragraph
2-2 for mechanical installation information.

b. Verify Terminal Block Wiring

Ensure the wiring of both the oxygen probe
terminal block and Model 3081 Transmitter
terminal block is correct. Refer to paragraph
2-3 for electrical installation and wiring
information.

3-2

POWER UP

a. General

The Two-Wire In Situ Oxygen Analyzer dis-
plays the current oxygen reading on the
LCD face of the Model 3081 Transmitter.
The O

2

concentration, cell temperature, and

4-20 mA output current are displayed as
shown in Figure 3-1. This and other infor-
mation may also be accessed using
HART/AMS.

b. Startup

Display

When the probe is first inserted into the
stack, some time is required until minimum
operating temperatures [550

°C (1022°F)]

are reached. Some time is also required for
the electronics to reach an operating state.
Therefore, when the unit is first powered up,
a faulted operation display as shown in Fig-
ure 3-2 may be displayed by the transmitter
until the probe operating temperatures are
reached and the electronics are working
properly (approximately 5 minutes).

c. Operating

Display

After the probe has reached operating
temperatures, the Model 3081 Transmitter
display should look similar to Figure 3-1.
The display will now track the O

2

concentra-

tion, cell temperature, and 4-20 mA output
current.

%

mA

4-20 mA OUTPUT

26020007

O CONCENTRATION

2

CELL

TEMPERATURE

Figure 3-1. Normal Operation Display

%

mA

26020008

Figure 3-2. Faulted Operation Display

3