Model 3081fg, Instruction manual – Emerson 3081FG User Manual

Page 15

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-5

Model 3081FG

TWO-WIRE IN SITU

OXYGEN ANALYZER

TERMINATION IN

CONTROL ROOM

24 VDC

POWER

SUPPLY

+

INTRINSIC

SAFETY

BARRIER

(OPTIONAL)

26020037

ASSET MANAGEMENT

SOLUTIONS

4-20 mA OUTPUT

(TWISTED PAIR)

HART MODEL 275

HAND HELD

INTERFACE

MODEL 3081

TRANSMITTER

O mV

SIGNAL

2

TEMPERATURE

mV SIGNAL

REFERENCE

AIR LINE

CALIBRATION CHECK

GAS LINE

Figure 1-2. Two-Wire In Situ Oxygen Analyzer HART Connections and AMS Application

e. Handling

the

Analyzer

The probe was specially packaged to pre-
vent breakage due to handling. Do not re-
move the padding material from the probe
until immediately before installation.

It is important that printed circuit
boards and integrated circuits are
handled only when adequate antistatic
precautions have been taken to pre-
vent possible equipment damage.

The oxygen probe is designed for in-
dustrial applications. Treat with care to
avoid physical damage. The probe
contains components made from ce-
ramic, which are susceptible to shock
when mishandled. THE WARRANTY
DOES NOT COVER DAMAGE FROM
MISHANDLING.

f. System

Considerations

Prior to installing your Rosemount Two-Wire
In Situ Oxygen Analyzer, make sure you
have all the components necessary to make
the system installation. Ensure all the com-
ponents are properly integrated to make the
system functional.

After verifying that you have all the compo-
nents, select mounting locations and deter-
mine how each component will be placed in
terms of available line voltage, ambient
temperatures, environmental considera-
tions, convenience, and serviceability. Fig-
ure 1-2 shows a typical system wiring. A
typical system installation is illustrated in
Figure 1-3.

1

Advertising