Hart communicator pc connections, Off-line and on-line operations, Hart communicator pc connections -3 – Emerson 3081FG User Manual

Page 49: Off-line and on-line operations -3, Two-wire in situ oxygen analyzer applications -3, Model 3081fg, Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

HART/AMS 4-3

Model 3081FG

LOOP CONNECTORS

USE INTERFACE

00275 0013 ONLY

SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS

SERIAL PORT

LOOP CONNECTORS

4-20 mA SIGNAL LINE

HART

COMMUNICATOR

RL < 250

ANALOG OUTPUT DEVICE

250 OHM
LOAD
RESISTOR
(SEE NOTE)

HART COMMUNICATOR

REAR PANEL

NOTE: THE SIGNAL LOOP MUST BE BROKEN

TO INSERT THE OPTIONAL 250 OHM
LOAD RESISTOR.

MODEL 3081 TRANSMITTER

TERMINAL BLOCK

12

3

5

6

7

8

9

11

12

13

14

15
16

10

4

+

4-20 mA

26020041

Figure 4-2. Signal Line Connections, < 250 Ohms Lead Resistance

4-3

HART COMMUNICATOR PC
CONNECTIONS

There is an option to interface the HART Com-
municator with a personal computer. Load the
designated AMS software into the PC. Link the
HART Communicator to the PC using the inter-
face PC adapter that connects to the serial port
(on the communicator rear panel).

Refer to the proper HART Communicator docu-
mentation in regard to the PC interface option.

4-4

OFF-LINE AND ON-LINE OPERATIONS

The HART Communicator can be operated both
off-line and on-line.

a. Off-line operations are those in which the

communicator is not connected to the Two-
Wire In Situ Oxygen Analyzer. Off-line op-
erations can include interfacing the HART
Communicator with a PC. (Refer to applica-
ble HART documentation regarding
HART/PC applications.)

b. In the on-line mode, the communicator is

connected to the 4-20 mA analog output
signal line. The communicator is connected
in parallel to the Two-Wire In Situ Oxygen
Analyzer or in parallel to the 250 ohm load
resistor.

c. The opening menu displayed on the HART

LCD is different for on-line and off-line op-
erations. When powering up a disconnected
(off-line) communicator, the HART LCD will
display the Main Menu. When powering up
a connected (on-line) communicator, the
HART LCD will display the On-line Menu.
Refer to the HART Communicator manual
for detailed menu information.

4-5

MENU TREE FOR HART COMMUNICATOR/
TWO-WIRE IN SITU OXYGEN ANALYZER
APPLICATIONS

This section consists of a menu tree for the
HART Communicator (Figure 4-3). This menu is
specific to Two-Wire In Situ Oxygen Analyzer
applications.

4