Identifying and correcting fault indications, Identifying and correcting fault indications -3, Model 3081fg – Emerson 3081FG User Manual

Page 61: Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 6-3

Model 3081FG

6-4

IDENTIFYING AND CORRECTING FAULT
INDICATIONS

A fault in the operation of the Two-Wire In Situ
Oxygen Analyzer is indicated by the faulted op-
eration display. If no faults exist, the display will
indicate NONE. Information on the current fault
is found under the DIAGNOSTICS MENU as
detailed in Section 3, STARTUP AND
OPERATION.

The following paragraphs describe the faults,
possible causes, and corrective actions. Refer
to Figure 6-4 as needed for test points and wir-
ing information.

NOTE

Allow adequate time for the oxygen
probe to reach its operating tempera-
ture [approximately 500

°°°°C (932°°°°F)] be-

fore investigating a fault. The SHOW
FAULTS screen of the DIAGNOSTICS
menu will indicate a fault until the unit
reaches operating temperature.

NOTE

The probe uses a Type B thermocou-
ple to measure the cell temperature. A
Type B thermocouple output table may
be useful for troubleshooting.

4-20 mA (+)

4-20 mA (- )

CELL AND

THERMOCOUPLE

4-20 mA

26020023

GROUND

THERMOCOUPLE

+

(GY)

CELL

-(

WH)

THERMOCOUPLE

-(

RD)

CELL

+

(BK)

EARTH

GROUND

TERMINALS

TERMINAL

BLOCK (TB1)

FACTORY-INSTALLED

JUMPER

PROBE CABLE

SHIELD