Troubleshooting, General, Probe life – Emerson 3081FG User Manual

Page 59: 0 troubleshooting -1 6-1, General -1, Probe life -1, Figure 6-1. slope vs. impedance -1, Model 3081fg, Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 6-1

Model 3081FG

SECTION 6

TROUBLESHOOTING

Install all protective equipment covers
and safety ground leads after trouble-
shooting. Failure to install covers and
ground leads could result in serious
injury or death.

6-1 GENERAL

This troubleshooting section describes how to
identify and isolate faults that may develop in
the Two-Wire In Situ Oxygen Analyzer.

6-2

PROBE LIFE

The zirconium oxide technology for measuring
oxygen is very stable and should provide accu-
rate service for several years.

Life of the probe is negatively impacted by:

• Continued operation at elevated tempera-

tures above 1300°C (2372°F).

• Operation in processes that contain high

levels of sulfur, SO

2

, or other acidic com-

pounds.

Operating conditions with simultaneously high
levels of SO

2

and low levels of O

2

are particu-

larly damaging.

The health and accuracy of a given cell is
closely related to the resistance, or impedance,
of the cell. Figure 6-1 illustrates that the amount
of output from a cell for a given O

2

value (repre-

sented as slope) will remain very stable to the
point where cell impedance increases to ap-
proximately 100 ohms.

SLOPE

(mV/Dec)

IMPEDANCE,

NORMAL

OPERATIONAL

CONDITIONS

48.0<S<51

mV/Dec

45.0<S<48

mV/Dec

S<45

mV/Dec

DETERIORATED PERFORMANCE,

SLOW RESPONSE,

HIGH MEASUREMENT ERROR

0

26020059

10

20

30

40

50

DETERIORATED PERFORMANCE,

SLOW RESPONSE,

HIGH MEASUREMENT ERROR.

CONDUCT FREQUENT CALIBRA-

TION CHECKS.

Figure 6-1. Slope vs. Impedance

6

Advertising