Startup and Operation 3-15

Model 3081FG

%

%

O CONCENTRATION

2

LOW CALIBRATION

CHECK GAS O VALUE

2

26020056

Figure 3-29. ACCEPT LOW O

2

d. SLOPE

Refer to Figure 3-30. Use this screen to ex-
amine the slope calculated from current
calibration check. The slope is the amount
of cell voltage generated for a given O

2

value. After each calibration check, record
the slope over the life of the probe. Tracking
the slope will indicate if the probe is de-
grading. Press NEXT to access the CON-
STANT screen.

%

O CONCENTRATION

2

CALCHECK CELL

SLOPE VALUE

26020057

Figure 3-30. SLOPE

e. CONSTANT

Refer to Figure 3-31. Use this screen to ex-
amine the cell zero constant calculated from
the current calibration check. The constant
represents the voltage generated by the cell
when no difference exists between the
amount of O

2

on the reference and process

sides of the cell. Note this value for com-
parison against future calibration checks.
Press RESET or EXIT to return to the
PROCESS DISPLAY screen.

%

O CONCENTRATION

2

CALCHECK CELL

CONSTANT VALUE

26020058

Figure 3-31. CONSTANT