Pneumatic installation, Pneumatic installation -11, Model 3081fg – Emerson 3081FG User Manual

Page 29: Instruction manual

Advertising
background image

Instruction Manual

IB-106-3081 Rev. 1.5

September 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-11

Model 3081FG

4-20 mA (+)

4-20 mA (-)

CELL AND

THERMOCOUPLE

4-20 mA

26020005

GROUND

THERMOCOUPLE

+

(GY)

CELL

-(

WH)

THERMOCOUPLE

-(

RD)

CELL

+

(BK)

EARTH

GROUND

TERMINALS

TERMINAL

BLOCK (TB1)

FACTORY-INSTALLED

JUMPER

PROBE CABLE

SHIELD

CONDUITS

NOTE: RUN CELL AND THERMOCOUPLE SIGNALS IN

SEPARATE CONDUIT FROM 4-20 mA LINE.

12

3

5

6

7

8

9

11

12

13

14

15

16

10

4

Figure 2-10. Model 3081 Transmitter Terminal

Block

2-4

PNEUMATIC INSTALLATION

a. General

Reference air is required for O

2

calculation,

and calibration check gas is required during
a calibration check. Refer to Figure 2-11 for
the gas connections on the oxygen probe.

b. Reference

Air

Package

After the oxygen probe is installed, connect
the reference air set. Install the reference air
set according to Figure 2-12.

c. Instrument Air (Reference Air)

Instrument air is required for reference. Use
10 psig (68.95 kPa gage) minimum, 225
psig (1551.38 kPa gage) at 0.2 scfh (100
ml/min.); less than 40 parts-per-million total
hydrocarbons. Regulator outlet pressure
should be set at 5 psi (35 kPa).

d. Calibration Check Gas

Two calibration check gas concentrations
are used with the Two-Wire In Situ Oxygen
Analyzer: Low Gas - 0.4% O

2

and High Gas

- 8% O

2

, each with the balance in nitrogen.

Do not use 100% nitrogen. See Figure 2-11
for the probe connections. Set both calibra-
tion check gases at the same flow rate: 5
scfh (2.5 L/min).

1/4 TUBE FITTING

(REFERENCE AIR PORT)

26020006

1/4 TUBE FITTING

(CALIBRATION CHECK

GAS PORT)

REFERENCE

AIR VENT

Figure 2-11. Oxygen Probe Gas Connections