Advanced Operations

6.3.3.6 Function set to PWM CONTROL/LINEAR working mode

PWM CONTROL/LINEAR

SET FUNCTION

Default value: LINEAR
The 2nd SET FUNCTION menu sets the PWM signal status

LINEAR

Signal is linearly dependent on the process value.

PID CONTROL

Signal works as a PID controller.

SET TRANSFER

Default value: 0 seconds
Sets a substitute PWM ratio when some error conditions are detected in the selected source or if
the source is missing.

HIGH ALARM

Default value: 15
Sets the process value which leads the PWM ratio to 100% (DUTY CYCLE set to DIRECT).

LOW ALARM

Default value: 5
Sets the process value which leads the PWM ratio to 0% (DUTY CYCLE set to DIRECT).

PERIOD (0 s

600 s)

Default value: 5 seconds
Sets the duration time of one PWM period.

MINIMUM
(0%

100%)

Default value: 0%
Lower limit of the operation range.

MAXIMUM
(0%

100%)

Default value: 100%
Upper limit of the operation range (

Figure 61

).

DUTY CYCLE

Default value: DIRECT
Sets the status of PWM ratio.

DIRECT

PWM ratio rises with a rising process value.

REVERSE

PWM ratio falls with a rising process value.

INPUT VALUE

Displays the process value read from the selected source after it has been processed by the
internal formula parser (if prompted).