A.4 Attach the external relay module

The external relay card provides 4 relays, each with one change-over contact. The relays
can switch maximum 250 VAC, 5A (UL, SPDT-CO, change-over). They can be
programmed for limits, status, timers, and special functions.

1. Remove power and disconnect all cable connections to the module.

2. Place the external relay module on the DIN rail to the right of the base module and

slide firmly against the base module (or other connected modules).

3. Make the appropriate connections as shown in

Figure 77

and

Table 50

.

4. Connect cables to the module and reconnect power from the instrument.

Figure 77 External relay module

Table 50 External relay module terminal assignments

Terminal

Assignment

Description

1

Normally opened contact 1/3 (NO)

Maximum switching voltage:

250 VAC;

125 VDC

Maximum switching current:

250 VAC, 5A
125 VAC, 5A

30 VDC, 5A

Maximum switching power:

1500 VA

150 W

2

Normally closed contact 2/3 (NC)

3

Not switched (COM)

4

Not used

5

Normally opened contact 5/7 (NO)

6

Normally closed contact 6/7 (NC)

7

Not switched (COM)

8

Not used

9

Normally opened contact 9/10 (NO)

10

Not switched (COM)

11

Normally closed contact 11/10 (NC)

12

Not used

13

Normally opened contact 13/14 (NO)

14

Not switched (COM)

15

Normally closed contact 15/14 (NC)

16

Not used